Operation – Leica Biosystems ASP6025 User Manual
Page 133

131
Leica ASP6025
Pressing the RUN OPTIONS button opens
the window with the same name.
Fig. 192
Starting a program
(continued)
• PRE-TEST activated:
After the program is started, the retort is filled and drained
with the first reagent in the infiltration program to test
whether all lines and valves are unobstructed.
• PRINTER, if activated:
At the end of the program, a complete run log covering the
entire program execution is stored in the form of a PDF file.
Fig. 191
If you start a pro-
gram for which TIME
REQUEST was selected
as the end time, an entry window (
Fig. 191
) opens in which you
can select the end time.
This function (
Fig. 192
) shows the options with which the
current program was started.
RUN OPTIONS
To ensure that the program ends on the speci-
fied date and time, the ASP6025 extends the
duration of the programmed delay step cor-
respondingly.
When working with fixed end times, you
should always check that the instrument time
and date are current.
The options set apply for ALL programs!
All run options can be enabled or disabled according to requirements while the program is running.
5. Operation