Cleaning and maintenance – Leica Biosystems ASP6025 User Manual
Page 153

151
Leica ASP6025
6. Cleaning and Maintenance
Running cleaning programs
(continued)
Fig. 213
Cleaning after completion of an infiltration program
If desired, you can select an additional water step for each retort
cleaning program (EXCEPT the full cleaning with detergents).
(See also
Chapter 6.1.3
).
Fig. 214
• To select the water step, press the WATER YES/NO button so
that a green check mark appears (
Fig. 212/213
). The water
step is always carried out as the third step in the cleaning
program, that is to say it is not run in addition to the cleaning
program shown in
Fig. 208
.
The three cleaning programs are offered au-
tomatically when an infiltration program has
been completed (
Fig. 211
).
In this case, a retort cleaning cycle MUST be
carried out; this function cannot be canceled.
The water step can only be selected if cleaning
water is present in one of the bottles.
In "Auto by RMS" mode, "Cleaning water"
must have been defined for a station.
In "Concentration" mode, the process water is
automatically used as cleaning water.
If the magnetic stirrer was removed from the
retort, it must be reinserted before the start
of cleaning.
Otherwise, the cleaning reagents in the retort
will not be heated effectively.
If cassettes/baskets are not removed immediately after a paraf-
fin run, paraffin can drop into the previously emptied retorts.
After an infiltration program, it is not possible
to avoid cleaning - no CANCEL key is present.
Cleaning MUST be carried out!
Fig. 212