Operation – Leica Biosystems ASP6025 User Manual
Page 98

96
Instructions for Use V 2.2 RevI – 08/2013
5. Operation
Deleting reagents from the list
• Highlight the reagent to be deleted in the SET UP REAGENTS AND WARNING THRESHOLDS
screen.
• Press DELETE.
• Press YES in the screen to confirm the reagent is deleted.
Fig. 134
Configuring the reagent list
(continued)
The ASP 6025 has a Reagent Management System (RMS) that ensures optimized reagent consump-
tion and the best possible tissue infiltration results.
The Reagent Management System is controlled via warning threshold values that display the
reagent replacement.
For each reagent, a value of this type can be defined depending on the use.
The RMS has two areas:
The warning threshold values, which are entered for all reagents in the reagent list and displayed
in the reagent status.
The reagent status also shows the values determined for the alcohols used (ethanol, isopropanol
and their mixtures with water).
The warning threshold values can be defined in the reagent list as:
• Number of processed cassettes
• Number of cycles (1 program sequence = 1 cycle) or
• Days until changing the reagent
The RMS is enabled by entering warning threshold values for one of these parameters specified
above in the reagent status.
5.2.2 The RMS system - adjusting warning threshold values
Please remember that a reagent which is already used in a program
cannot be deleted.