Instrument setup – Leica Biosystems ASP6025 User Manual
Page 51

49
Leica ASP6025
Pull-out drawer with RTU-bottles
(continued)
Height
(max.):
350 mm
Height
(min.):
245 mm
Width
(max.):
200 mm
Width
(min.):
155 mm
Depth
(max.):
180 mm
Depth
(min.):
135 mm
4. Instrument Setup
Permitted dimensions for RTU-bottles in the drawer:
Bottle neck diameter
Outer (
max.
): 54 mm
Outer (
min.
): 38 mm
Inner (
max.
): 44 mm
Inner (
min.
):
27 mm
• The RTU-bottles are displayed in the SMART SCREEN in a
separate area on the screen. They are designated "D1" to
"D6" (
Fig. 48
).
• Additionally, the RTU-bottles can be emptied using the
remote drain system.
• The hose for the remote fill and drain system can also be
stored in the drawer.
Fig. 48
If RTU-bottles other than those installed by Leica are used in the drawer, these must be resistant to the
following reagents and temperatures and dimensionally stable.
For cleaning solutions up to a temperature of 71 °C
For progress reagents up to a temperature of 64 °C
The dimensions must not fall below or exceed those listed below.