Operation – Leica Biosystems ASP6025 User Manual
Page 108

106
Instructions for Use V 2.2 RevI – 08/2013
5. Operation
In the following program (that is to say when another infiltra-
tion program is started in Concentration mode), the ethanol is
replaced.
For this purpose, fresh ethanol must be added manually to
RTU-bottle D4 in the drawer prior to the start of the program.
The operator is prompted to do this by a corresponding screen
instruction (
Fig. 146
).
This must be confirmed with YES before the program is started.
The ethanol is then transferred to system bottle S6 while the
program is running.
Functional description of ethanol rotation
(continued)
If NO (
Fig. 146
) is pressed, the adjacent message is displayed.
Pressing YES (
Fig. 148
) starts the program - WITHOUT ethanol
rotation! - Refilling system bottle S6 is done manually and must
be confirmed in the reagent status (
Fig. 140
, p. 101).
Pressing NO stops the program or is stopped or does not start it.
Fig. 147
Fig. 146
Fig.
148
2. step
Pressing OK (
Fig. 147
) starts the program - fresh ethanol is used.