Measuring the alcohol concentration
The MEASURE CONC. field (
Fig. 85
) shows whether or not alco-
hol concentration measurement is enabled or not.
In concentration mode (
Fig. 79
), alcohol concentration measure-
ment is ALWAYS enabled.
Alcohol concentration can also be measured in RMS mode. If this
is not desired, it can be disabled.
Press the button for the setting desired in the
MEASURE CONC.
field (Fig. 85/86). A green check appears for the selected option.
5. Operation
Fig. 85
Fig. 86
Fig. 87
When alcohol concentration mea-
surement is enabled, it takes place
within a cleaning program. This
increases the time of the cleaning
by a few minutes.
The measured concentrations ap-
pear in the REAGENT STATUS under
CONCENTRATION FOR ALCOHOL
(
Fig. 87
).
The value configured for THRESHOLD VALUE/CONCENTRATION has no influence on the Reagent Management
System (RMS) - there is no warning value! In this case, the measured concentrations are only informative.