Troubleshooting, 1 troubleshooting, 1 power failures – Leica Biosystems ASP6025 User Manual
Page 167: 1 troubleshooting 7.1.1 power failures

165
Leica ASP6025
7. Troubleshooting
7.1 Troubleshooting
When a problem occurs in the Leica ASP6025, take the following steps to diagnose the problem:
•		 Check	for	error	messages	indicating	the	nature	of	the	problem.
•		 If	an	error	message	is	displayed,	press	the	Help	button	on	the	message	for	any	information	
relating to the cause.
• Check the Run Log for any messages that indicate the nature of the failure. If the failure
 occurred during a program run, check the Run Log for an indication of whether the failure 
occurred during the fill, processing or drain cycles, and the station in use at the time.
7.1.1 Power failures
If there is evidence of power to the instrument:
•		 Verify	that	the	power	plug	is	plugged	into	the	socket	and	that	the	socket	is	switched	ON.	
•		 Verify	that	the	power	switches	on	the	rear	of	the	instrument	(adjacent	to	the	power	lead),	and	
on the side of the instrument are both switched ON.
• Error messages indicating that there has been a power failure mean that power has been lost
to the heaters and the monitor, but not to the main control electronics. Perform test procedure 
as described in "Power failure" (
Chapter 7.1.2
).
Use the functions in the SMART SCREEN menu to test operating steps individu-
ally, such as the fill, drain, pressure, and vacuum steps.
If you cannot identify the problem by carrying out the above listed steps, call
Leica Technical Service.