2 adding and modifying programs, Operation – Leica Biosystems ASP6025 User Manual
Page 112

110
Instructions for Use V 2.2 RevI – 08/2013
Creating a new program
• Make sure you are logged on at supervisor level.
• In the VIEW/EDIT USER PROGRAMS menu (
Fig. 152
) highlight a program as similar as pos-
sible to the program you wish to create - this minimizes the number of steps that are then
necessary to create it).
• Press COPY to copy the selected program. The new program will have the same name as the
program copied, however, the figure "(2)" will be added to indicate the change.
• Highlight the line containing the new program.
• Touch the PROGRAM NAME button at the top of the table, the keyboard will appear (
Fig. 153
).
• Enter the new program name.
Editing program steps
• Pressing the EDIT button (in
Fig. 152
) calls up the Program steps screen (
Fig. 154
).
• In the headline you will find the PROGRAM NAME.
• The colors on the left border of the table indicate the reagent groups to which the reagents
belong.
• The program steps are displayed in the order in which they are carried out. For each program,
up to 15 steps can be defined.
The following characteristics of each program step can be edited:
(
Fig. 154
)
• The reagent that is used in the step
• Duration of step
(not including the filling and emptying times)
• The retort temperature
(if "Ambient" is selected, the display for the retort temperature remains empty)
• Type of pressure and/or vacuum cycle
• Set the delay step.
Fig. 153
5.3.2 Adding and modifying programs