Troubleshooting – Leica Biosystems ASP6025 User Manual
Page 169

167
Leica ASP6025
7. Troubleshooting
• If no infiltration program is running when a power failure is detected, no further actions / backup measures are taken -
the instrument shuts down automatically.
When a program is running:
• The program that is running is stopped, regardless of the current status of processing.
Fig. 239
• If a reagent is being used for which no safe reagent is in the
instrument when the program starts, a warning message
appears (
Fig. 239
). However, the program launch is NOT
aborted.
• If a safe reagent is in the retort, the reagent can be kept in
the event of power failure, though the retort may have to be
filled to a high enough level so that no tissue sample is dry.
This operation is automatically controlled by the instrument
and cannot be influenced by the operator.
• If no safe reagent is in the retort, depending on the cur-
rent status of processing the program that is in progress, a
"decision" is made about how to proceed.
Current step for power failure
Safeguard in safe reagent
Formalin (fixation)
Formalin
Ethanol (dehydration, diluted)
Formalin
Ethanol (absolute)
Formalin
Xylene (intermedium)
Xylene (heating off)
Isopropanol (intermedium)
Formalin
Paraffin
Paraffin (heating off)
ParaLast ParaLast
To conserve the battery, all heaters, the agitator and the monitor are switched off directly.