6. Cleaning and Maintenance
Replacing the active carbon filter
• The life of the active carbon filter will depend on the reagent
types used and how frequently they are used in the instru-
ment.
• The filter (13) should be replaced at least every 30 days
(
Fig. 227
).
After replacing the filter, open the SYSTEM MONITOR menu
to set the filter replacement date to the current date. See also
Chapter 5.1.4.
Fig. 227
Emptying the condensate bottle
• The blue condensate bottle must be taken out, emptied and
reinserted at regular intervals (weekly). This should be done
sooner if a prompt to that effect is issued by the software
(
Fig. 225
).
Fig. 225
Filter label
14
13
Fig. 226
Dispose of waste solvents with care according
to local regulations and the waste manage-
ment policy of the company or institution.
When inserting the new filter, make sure that
it is pushed all the way into the compartment
and that the flap is closed.