Operation – Leica Biosystems ASP6025 User Manual
Page 124

122
Instructions for Use V 2.2 RevI – 08/2013
5. Operation
Draining a paraffin bath
(continued)
Fig. 177
• The following security prompt must be confirmed in order
to continue (
Fig. 178
).
Fig. 178
• If necessary, the retort must be heated to the necessary
temperature so that the paraffin remains fluid. Then, the
paraffin is pumped from the selected paraffin bath into the
retort.
• When the bath is emptied, the paraffin is automatically
pumped from the retort into the external container.
• After this operation is complete, the paraffin residues are
blown out of the external tube and the hose. The correspond-
ing message must be confirmed with OK.
Fig. 179
+
Fig. 176
• Select the paraffin tank to be emptied and the receptacle for
the used paraffin (external station) (
Fig. 176
).
• Then press STATION DRAIN.
If the last reagent in the retort is not compatible with paraffin, the retort must first be cleaned using a cleaning
program.
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)