beautypg.com

Nanodac – Carbolite nanodac User Manual

Page 90

background image

nanodac™

MC27 –EN– 1.04

88

ANALOGUE INPUT SPECIFICATION (Cont.)
DC Input ranges (Cont.)

10V Range

Range:

-3V to +10V

Resolution

500μV

Measurement noise:

550μV peak-to-peak with 1.6s input filter

Linearity error:

0.007% (best fit straight line) for zero source resistance. Add 0.003% for each 10Ω source and lead resistance

Calibration error:

±1.5mV ±0.063% measurement at 25°C ambient

Temperature coefficient:

±66μV/°C ± 45ppm/°C of measurement from 25°C ambient

Input resistance:

62.5kΩ for input voltages > 5.6V. 667kΩ for input ranges < 5.6V.

Note: 10V range not available for dual input channels

Resistance input ranges

Temperature scale

ITS90

RTD Types, ranges and accuracies

See table

Maximum source current

200μA


Resistance input figures

Range:

0 to 400Ω (-200 to +850°C)

Resolution:

0.05°C

Measurement noise:

0.05°C peak-peak with 1.6s input filter

Linearity error:

0.0033% (best fit straight line)

Calibration error:

±0.31°C ±0.023% of measurement in °C at 25°C ambient

Temperature coefficient:

±0.01°C/°C ±25ppm/°C measurement in °C from 25°C ambient

Lead resistance

0 to 22Ω matched lead resistances

Bulb current:

200μA nominal













Table A3a RTD type details

Thermocouple data

Temperature scale

ITS90

CJC

Types:

Off, internal, external, remote.

Remote CJC source

Any input channel

Internal CJC error

<1 °C max, with instrument at 25 °C

Internal CJC rejection ratio:

40:1 from 25°C

Upscale/downscale drive

High, low or none independently configurable for each channel’s sensor break detection.

Types, ranges and accuracies

See table A3b



























Table A3b Thermocouple types, ranges and accuracies


Cu10
Cu53

JPT100

Ni100
Ni120

Pt100

Pt100A

RTD type

Overall range

°C

Standard

Max. linearisation error

-20 to + 400
-70 to + 200

-220 to + 630

-60 to + 250
-50 to + 170

-200 to + 850
-200 to + 600

General electric Co.

RC21-4-1966

JIS C1604:1989

DIN43760:1987
DIN43760:1987

IEC751

Eurotherm Recorders SA

0.02

°C

<0.01

°C

0.01

°C

0.01

°C

0.01

°C

0.01

°C

0.09

°C

C
D
E

G2

J

K
L

N

R
S
T
U

N iMo/ N iCo

Platinel

Mi/ NiMo

Pt20%Rh/Pt 40%/ Rh

0 to + 2300
0 to + 2495

-270 to + 1000

0 to + 2315

-210 to + 1200
-270 to + 1372
-200 to + 900
-270 to + 1300

-50 to + 1768
-50 to + 1768

-270 to + 400
-200 to + 600

-50 to + 1410

0 to + 1370
0 to + 1406
0 to + 1888

B

0 to + 1820

Hoskins
Hoskins

IEC584.1

Hoskins

IEC584.1
IEC584.1

DIN43710:1985 (to IPTS68)

IEC584.1
IEC584.1
IEC584.1
IEC584.1

DIN43710:1985
ASTM E1751-95

Engelhard

Ip sen

ASTM E1751-95

I EC584.1

0.12°C
0.08°C
0.03°C
0.07°C
0.02°C
0.04°C
0.02°C
0.04°C
0.04°C
0.04°C
0.02°C
0.08°C
0.06°C
0.02°C
0.14°C
0.07°C

0 to 400°C = 1.7°C

400 t o 1820°C = 0.03°C

T/ C t ype

O verall range (°C)

St andard

Max. linearisat ion error