To Set Up OEM Security
This will be illustrated by the following two examples:
EXAMPLE 1:
Make the parameter Network.Modbus.TimeFormat read/write when OEM security is enabled and the
instrument is in Engineer access level and leave the remaining parameters in this list as read only.
Select Engineer (Configuration)
access level
With OEM status unlocked set
‘OEMParamsLists’ to On.
Figure 3.6.10a
Open ‘Network.Modbus list
Figure 3.6.10b
Open ‘Instrument.OEMConfigList’
Drag and drop the parameter(s) which
are required to be read/write in
Engineer level when OEM security is
enabled.
In this example the parameter
‘TimeFormat’.
Figure 3.6.10c