1 “logger operation” mode – Dwyer PSX2 User Manual
Page 86
77
10.2.1 “Logger Operation” mode
There are two logging modes, i.e., quick logger that permits operation from the measurement
screen and logger that is set from the menu screen. Logger is of two different modes, i.e., con-
tinuous mode and appointed time mode.
• Quick logger
• Logger
(1) Continuous
mode
(2) Appointed time mode
(1) Quick logger
The quick logger starts when quick logger start button
in the measurement screen is pressed.
The quick logger exits when a period of an hour has elapsed since it was started or when quick
logger stop button is pressed.
• Logging time: 1 hour, fi xed
• Cycle: 10 seconds, fi xed
• Measured data type: 3 types (unit and number of digits after decimal point are the same) dis-
played in measurement screen and status display
(2) Logger
(1) Continuous mode
Continuous mode is the mode to perform logging in a fi xed period from start date and time to
exit date and time.
Exit occurs upon elapse of exit date and time or when the stop button is pressed. As the exit
time varies by the start time and period, there are cases where it is different from the set exit
time.
• Logging time: Start date and time to exit date and time
• Cycle: 10 seconds to 24 hours
• Measured data type: 14 measured data types and status display
Bulletin F-9-PUX2
Start button
Measurement screen