(3) Pipe is always fi lled with fl uid. Neither air bubbles nor foreign materials are contained in the

fl uid.

(4) There is an ample maintenance space around the pipe to which the detector is to be mounted

(see fi gure below).
Note 1) Secure an adequate space for allowing a person to stand and work on both sides of a

pipe.

Note 2) D indicates the inside diameter of a pipe.

D + 1200 or more

600

or more

D

600

or more

200

or more

200

or more

No

te

2000 or m

o

re

D : Pipe diameter

Space required for mounting detector

(5) The piping must completely be fi lled with fl uid when it fl ows.

Pump

Air-collecting

Good

Good

Pipe may
not be filled
with liquid.

Pipe may not be filled with liquid.

(6) For a horizontal pipe, mount the detector within ±45° of the horizontal plane.

For a vertical pipe, the detector can be mounted at any position on the outer circumference.

Horizontal plane

Pipe

Deposits of
sludge

Air bubbles

45°

45°

Bulletin F-9-PUX2

(47)

(23.6)

(23.6)

unit: mm (in.)