13. EXTERNAL COMMUNICATION SPECIFICATION
(1) General specifi cation
Item
Specifi cation
Transmission scheme
Half duplex
Synchronization scheme
Asynchronous
Transmission rate
500kBPS
Parity
Odd parity
Start/stop bit
1 bit
Data length
8 bits
Station
0, fi xed
Number of connectable units
1 unit
Transmission code
Hexadecimal value (MODBUS RTU mode)
Error detection
CRC-16
Echo back
None
Flow control
Xon/off
(2) Interface specifi cation
Electrical specifi cation: Conforms to USB standard.
Cable length: 3 m or less
Conforming cable: Mini USB cable
Connection scheme: 1:1 connection
Support software
Loader software for PC is provided as standard.
• Main function: To display and change main unit parameters (site setup), and to acquire mea-
sured data.
It is capable of importing instantaneous fl ow rate, instantaneous fl uid velocity, integrated
value, error information, wave receiving level and others.
Bulletin F-9-PUX2