beautypg.com

Dwyer PSX2 User Manual

Page 147

background image

138

(4) Fluid temperature

Point the cursor to “FLUID TEMPERATURE” by
using the

,

key and press the

ENT

key.

Input the fl uid temperature.
The status for numerical value input is produced
when the

ENT

key is pressed.

Move the cursor to the point to change the numerical
value, and change the numerical value using

,

keys.
Finalize the numerical value by pressing the

ENT

key.

(5) Measuring range

Point the cursor to “MEAS. RANGE” by using the

,

key and press the

ENT

key.

Select the range of fl ow velocity distribution to be
measured.
Radius F: Measurement is taken on the radius side

opposite to the sensor. (Normally used.)

Radius N: Measurement is taken on the radius side adjacent the mounted sensor.

Diameter: Measurement is taken across the entire diameter.

(6) Setup of judgment value

Selects a page by using

,

key.

Set values for judgment of whether fl ow velocity distribution measurement is normal or
abnormal. Success rate error arises, if measured values are less than judgment values. (Nor-
mally not used.)

Bulletin F-9-PUX2

This manual is related to the following products: