(4) Fluid temperature
Point the cursor to “FLUID TEMPERATURE” by
using the
,
key and press the
ENT
key.
Input the fl uid temperature.
The status for numerical value input is produced
when the
ENT
key is pressed.
Move the cursor to the point to change the numerical
value, and change the numerical value using
,
keys.
Finalize the numerical value by pressing the
ENT
key.
(5) Measuring range
Point the cursor to “MEAS. RANGE” by using the
,
key and press the
ENT
key.
Select the range of fl ow velocity distribution to be
measured.
Radius F: Measurement is taken on the radius side
opposite to the sensor. (Normally used.)
Radius N: Measurement is taken on the radius side adjacent the mounted sensor.
Diameter: Measurement is taken across the entire diameter.
(6) Setup of judgment value
Selects a page by using
,
key.
Set values for judgment of whether fl ow velocity distribution measurement is normal or
abnormal. Success rate error arises, if measured values are less than judgment values. (Nor-
mally not used.)