beautypg.com

Caution, 4 mounting of medium type sensor on pipe – Dwyer PSX2 User Manual

Page 60

background image

51

8.5.4 Mounting of medium type sensor on pipe

Mounting the detector using the following procedure.

(1) Provide wire rope for the upstream and the

downstream detectors.
Make sure that the length of the wire rope is
longer than the circumference of the pipe.

(2) Lay the wire rope around the pipe at the position

of the upstream detector.
Then hook the mounting spring into the wire
rope.

(3) Spread silicone fi ller over the whole transmitting

side of the sensor.
Care should be taken to prevent entry of air
bubbles.

(4) After cleaning the surface of the pipe, the sensor

should be mounted.

(5) Spread the wire rope near the marked lines in

the left-right direction, bring the sensor in close
contact and fi t the wire rope.
Make sure that the matching mark on the sensor
is aligned with the marking line.

Transmitting mark

Matching mark

Marking
line

Sensor

Marking line

Mounting spring

Pipe

CAUTION

Be careful not to cut your hands or etc. by the wire rope.

Bulletin F-9-PUX2

This manual is related to the following products: