8.5.4 Mounting of medium type sensor on pipe
Mounting the detector using the following procedure.
(1) Provide wire rope for the upstream and the
downstream detectors.
Make sure that the length of the wire rope is
longer than the circumference of the pipe.
(2) Lay the wire rope around the pipe at the position
of the upstream detector.
Then hook the mounting spring into the wire
rope.
(3) Spread silicone fi ller over the whole transmitting
side of the sensor.
Care should be taken to prevent entry of air
bubbles.
(4) After cleaning the surface of the pipe, the sensor
should be mounted.
(5) Spread the wire rope near the marked lines in
the left-right direction, bring the sensor in close
contact and fi t the wire rope.
Make sure that the matching mark on the sensor
is aligned with the marking line.
Transmitting mark
Matching mark
Marking
line
Sensor
Marking line
Mounting spring
Pipe
CAUTION
Be careful not to cut your hands or etc. by the wire rope.