Page 54
45
8.4.2 How to mount a small size (standard) sensor (Z method)
(1) Remove saddle set screws at 4 loca-
tions, and remove a saddle and a sen-
sor unit out of the frame.
Also, remove a saddle on the guide rail
for small size sensor (option).
Element holder
Lock nut
Saddle
Scale
Curser
Frame
BNC connector
Guide rail for small size sensor
(2) Mount the removed sensor unit on the
guide rail for small size sensor.
Fasten the sensor unit with mounting
dimension (L).
Mounting
dimension (L)
(3) Spread silicone grease over the whole
transmitting surface of the sensor.
Turn the element holder counterclock-
wise to return the sensor.
After cleaning the surface of the pipe,
the sensor should be mounted.
Element holder
Silicone grease
Transmitting surface
(4) Mount each sensor individually on the
marking line.
Front view
Back view
Upper side of
the marking line
Bottom side of
the marking line
Bulletin F-9-PUX2