beautypg.com

0 test voltage selection procedure, Test voltage selection procedure, Table 19.0 – Vanguard Tri-Phase User Manual

Page 111: Transformer test voltage selection procedure

background image

TRI-PHASE OPERATING™ INSTRUCTIONS

111


19.0 Test Voltage Selection Procedure

Table 19

.0 shows the procedure to perform in order to select a transformer test voltage. Refer to

section 19.1 for a description of the available transformer test voltages.

Table 19.0

Transformer Test Voltage Selection Procedure

STEP DESCRIPTION

DISPLAY

ACTION

1

Select “Setup” from “Main
Menu”

1.RUN TEST

.

2.SETUP
3.TEST PLAN
4.DIAGNOSTIC
TIME: 20:15:00
DATE: 07/16/08

Press key number 2
Selection of 2 may be
made by turning the
Control Knob. Push
down Control Knob
after 2 is selected

2

Select “Test Voltage” from the
“Setup Menu”

1.RECORD ID

.

2.TEST VOLTAGE
3.TEST FREQUENCY
4.PRINT RECORD
5.SAVE/RESTORE RECORD
6.SET TIME

Press key number 2
Selection of 2 may be
made by turning the
Control Knob. Push
down Control Knob
after 2 is selected

3 Select

“8

Volts”

1. 8 Volts

.

2. 40 Volts
3. 100 Volts

Press key number 1 or
push down Control
Knob
(8 volts is selected)

4

Confirm test voltage

8 VOLTS SET

Press any key or push
down Control Knob

5

Return to “Main Menu”

1.RUN TEST

.

2.SETUP
3.TEST PLAN
4.DIAGNOSTIC
TIME: 20:15:00
DATE: 07/16/08

None