Summit T3‐8 PCI Express Multi‐Lane Protocol Analyzer User Manual

21

Expansion Mode using Two Summit T3‐8 Analyzers

 Teledyne LeCroy

2 Multi‐lead Probe pods for x8
(

See “1 Multi‐lead Probe Pod for x1 and x4” on page 19

).

Up to 16 MidBus Probe SSMP Cables, and up to 8 Flexible Leads, per pod.
(

See Figure 3.17 on page 19

).

1 Clock Cable (

see

 

Figure 3.15 on page 18

).

Connections Overview for Gen2 Multi-lead Probe

Use a 1‐pod setup.

For x8, use the straight iPass cables to connect the probe data connectors on the Analyzer 
to the pod(s). 

On the other side of the pod, connect the flexible lead tips.

Connection Procedure

To connect Summit T3‐8 to the System Under Test (host platform or root complex under 
test): 

1. Connect the pods to the Analyzer using the straight iPass cables.
2. Connect the flexible lead tips to the pods using the SSMP cables.
3. Connect the lead tips to the system under test by soldering to the trace. 

3.10 Expansion Mode using Two Summit T3-8 Analyzers

Expansion Mode

You can setup two Summit T3‐8 analyzers to work together as one analyzer, in Expansion 
Mode, to support x16 recording capability. 

To set up the hardware configuration, both analyzers must have a CATC‐Sync card plugged 
into a front or back expansion slot. Then perform the following steps:

1. Before setting up the hardware configuration, turn power off on both analyzers.
2. Using a CATC‐Sync Cable, connect the Sync‐Out connector on the CATC‐Sync card of 

one Summit T3‐8 analyzer to the Sync‐In connector on the CATC‐Sync card of the 
second Summit T3‐8 analyzer.

3. Using an x16 Expansion Cable, connect the Expansion connector on the back of one 

Summit T3‐8 analyzer to the Expansion connector on the back of the other 
Summit T3‐8 analyzer.

4. Connect both analyzers to the host machine that has the PETracer application.
5. After completing the hardware configuration, power on both Summit T3‐8 

analyzers.

6. Open the PETracer application on the host machine.

The software automatically detects and configures the analyzers to work as one x16 
‐ capable analyzer. One Summit T3‐8 analyzer is assigned the Upstream direction, 
and the other analyzer is assigned the Downstream direction. The LCD display on the 
front of each analyzer displays its direction assignment.