Teledyne LeCroy WavePro Automation Command User Manual
Page 521

Automation Command and Query Reference Manual - Processor Reference
Enables/Disables automatic scale setting for the histogram function Fx.
Description
Bins
DoubleLockstep
From 20 to 2000 step 1, locked to 1 2 5, fine grain allowed=false, on=false
Range
Sets/Queries the number of bins in the histogram function Fx.
Description
Center
Double
From -1e+010 to 1e+010 step 1e-012
Range
Sets/Queries the horizontal value at the center of the graticule of the histogram function Fx.
Description
ClearSweeps
Action
Clears the contents of all the bins of the histogram function Fx.
Description
FindScale
Action
Creates a suitable horizontal position and scale to include all the non-empty bins of the histogram Fx.
Description
HorScale
DoubleLockstep
From 1e-012 to 1e+012 step 0.01, locked to 1 2 5, fine grain allowed=false, on=false
Range
Sets/Queries the horizontal scale in units per division for the histogram function Fx. Use the FindScale
control to automatically determine the scale by looking at the non-zero populated bins.
Description
Values
Integer
From 20 to 2000000000 step 1
Range
Sets/Queries the maximum number of values from the source result to include in the histogram
function Fx.
Description
2-22