Teledyne LeCroy WavePro Automation Command User Manual
Page 208

Automation Command and Query Reference Manual - Control Reference
Clear4
Action
Clear all calibration coefficients for Channel 4.
Description
CommonMode
Double
From -1.1 to 1.1 step 0.001
Range
The TF-DSQ fixture will calibrate probes differentially or in single-ended mode depending on the type
of probe. Differential probes allow the common mode voltage component to be applied during the DC
calibration for improved calibration accuracy in situations where probe gain or offset correction
depends on common mode components.
Description
DcCal1
Action
Initiate a DC calibration (Gain/Offset) of Channel 1.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
DcCal2
Action
Initiate a DC calibration (Gain/Offset) of Channel 2.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
DcCal3
Action
Initiate a DC calibration (Gain/Offset) of Channel 3.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
DcCal4
Action
Initiate a DC calibration (Gain/Offset) of Channel 4.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
Deskew1
Action
Initiate a deskew calibration of Channel 1.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
Deskew2
Action
Initiate a deskew calibration of Channel 2.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
Deskew3
Action
Initiate a deskew calibration of Channel 3.
(Note that this action pops up an interactive wizard, which requires a user to manually dismiss)
Description
1-194
918501 RevA