Teledyne LeCroy WavePro Automation Command User Manual
Page 210

Automation Command and Query Reference Manual - Control Reference
ProbeName1
String
Any number of characters
Range
Reflects the name of the probe detected on Channel 1. Contains the text 'No Probe', if no probe is
sensed.
Description
ProbeName2
String
Any number of characters
Range
Reflects the name of the probe detected on Channel 2. Contains the text 'No Probe', if no probe is
sensed.
Description
ProbeName3
String
Any number of characters
Range
Reflects the name of the probe detected on Channel 3. Contains the text 'No Probe', if no probe is
sensed.
Description
ProbeName4
String
Any number of characters
Range
Reflects the name of the probe detected on Channel 4. Contains the text 'No Probe', if no probe is
sensed.
Description
ProbeName5
String
Any number of characters
Range
RecallCalSetup
Action
Whenever a probe calibration is applied, the scope saves the information in a file on the disk. If the
scope must be rebooted for any reason, the probe calibration information is always cleared, but can
be recalled using this control.
Description
RiseTime1
Double
From 0 to 2e-009 step 1e-012
Range
When probes are deskewed, this control is set to the measured risetime of Channel 1.
Description
RiseTime2
Double
From 0 to 2e-009 step 1e-012
Range
When probes are deskewed, this control is set to the measured risetime of Channel 2.
Description
1-196
918501 RevA