Result, Vtxcm, Vtxde – Teledyne LeCroy WavePro Automation Command User Manual
Page 372
Automation Command and Query Reference Manual - Control Reference
View
Bool
Sets/Queries the trace's 'Viewed' state. When true, the trace is displayed on one of the display
graticules. Note that even when a trace is not visible, it may be used as a source for Math, Measure,
etc.
Description
RESULT
app.SDA2.PCIe.Parameters.VdiffPP.Out.Result
VTXCM
app.SDA2.PCIe.Parameters.VTxCm
Transmitter common mode voltage
ClearSweeps
Action
Clear any accumulated result data. Useful for example to restart an average, or parameter statistics.
Description
View
Bool
Sets/Queries the trace's 'Viewed' state. When true, the trace is displayed on one of the display
graticules. Note that even when a trace is not visible, it may be used as a source for Math, Measure,
etc.
Description
ClearSweeps
Action
View
Bool
RESULT
app.SDA2.PCIe.Parameters.VTxCm.Out.Result
VTXDE
app.SDA2.PCIe.Parameters.VTxDe
De-emphasis in dB. This measurement should be made at the transmitter's pins, otherwise what is being measured
is the remaining de-emphasis after some is used up by the fixture or channel.
ClearSweeps
Action
Clear any accumulated result data. Useful for example to restart an average, or parameter statistics.
Description
ClearSweeps
Action
View
Bool
1-358