Plltrack – Teledyne LeCroy WavePro Automation Command User Manual
Page 351
Automation Command and Query Reference Manual - Control Reference
ShowRjBUjTrack
Bool
When this control is set to True the RjBUj Track is displayed. The RjBUj track is a time domain
waveform spanning the same time as the input waveform, and with vertical units of time. It shows the
random jitter and the bounded uncorrelated jitter (but not the DDJ, which is correlated to the data
pattern). Compare this to the TIE track, which includes all the jitter. They may look visually the same,
but chances are the TIE track is at a larger vertical scale. TIE track is always greater than or equal to
the RjBUj track.
Description
ShowTIETrack
Bool
When this control is set to True the TIE Track is displayed. TIE Track is a time domain waveform,
spanning (almost) the same time as the input waveform, and with vertical units of time. TIE Track
shows all the jitter that was in the input signal. Compare it to RjBUj track, see ShowRjBUjTrack,
above.
Description
PLLTRACK
app.SDA2.JitterMeasure.Track.PLLTrack
AxisXRotation
Integer
From -90 to 90 step 1
Range
AxisYRotation
Integer
From -90 to 90 step 1
Range
This control is used only when Persisted is true and Persistence3d is true. It controls rotation about
the Y axis of the view being persisted.
Description
AxisXRotation
Integer
AxisYRotation
Integer
ClearSweeps
Action
LabelsPosition
String
LabelsText
String
Persist3DQuality
Enum
PersistDotJoined
Bool
Persisted
Bool
Persistence3d
Bool
PersistenceMonoChrome
Bool
PersistenceSaturation
Integer
PersistenceTime
Enum
ShowLastTrace
Bool
UseGrid
String
View
Bool
ViewLabels
Bool
1-337