84

H-Class

Advanced Entry Calibration (continued)

Step

Action

Displayed Message

Comment

F

Use the buttons to set the

Gain Number to 00 and
then press ENTER.


Record the sensor reading

as a Label Value for Gain
Number 00 in a table (32
rows by four columns,

with headings similar to
those shown below.)

TRAN SENSOR GAIN


yyy (0 - 31)

00

This is the Label Value for

a gain setting of 00,
where “yyy” represents

the current sensor
reading.


Sample Calibration Table

Gain Number

Label Value

TOF Value

Difference Value

00 252

01

02

31



Step

Action

Displayed Message

Comment

G

Use the buttons to
increment the Gain

Number by one and then
press ENTER. Record the
Label Value. Repeat this

process for each Gain
Number.

TRAN SENSOR GAIN

yyy (0 - 31)
01

These are Label Values,
where “yyy” represents

the current sensor
reading.


Sample Calibration Table

Gain Number

Label Value

TOF Value

Difference Value

00 252

01 250

02 248

… …

31 09