beautypg.com

Tektronix AWG610 User Manual

Page 422

background image

Appendix B: Performance Verification

B-62

AWG610 Arbitrary Waveform Generator User Manual

3. Set the oscilloscope controls as follows:

Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1 and CH2

CH1 and CH2 coupling . . . . . . . . . DC
CH1 and CH2 scale . . . . . . . . . . . . 1 V/div
CH1 and CH2 offset. . . . . . . . . . . 0 V
CH1 and CH2 inputimpedance . . . 50 W

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 500 ps/div

Trigger

Source . . . . . . . . . . . . . . . . . . . . . CH2
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Normal

4. Follow the substeps below to set the AWG610 Arbitrary Waveform

Generator controls and select the waveform file:

a. Push UTILITY (front-panel)

!System (bottom)!Factory Reset

(side)

!OK (side).

b. Load the MODE.WFM file.

Refer to Loading Files on page B–10 for file loading procedures.

5. Push the RUN button.

The LED above the RUN button is on.

6. Follow the substeps below to verify that the Marker1 delay function is

operating correctly:

a. Push HORIZONTAL MENU (front-panel)

!Marker1 Delay (side)

b. Continuously change the Marker1 delay from 0 s to 1.5 ns by turning the

general purpose knob.

c.

Verify that the Marker1 output delay is relative to the Marker2 output
and varies from 0 s to 1.5 ns on the oscilloscope screen.

d. Verify that the Marker1 output delay relative to the Marker2 output is

in 1.05 ns to 1.65 ns at marker1 delay setting is 1.5 ns.

7. Push 0 and ENTER keys on the AWG610 Arbitrary Waveform Generator to

set the Marker1 Delay back to 0 s.