Tektronix AWG610 User Manual
Page 173
The Graphical Waveform Editor
AWG610 Arbitrary Waveform Generator User Manual
3-55
Do the following steps to vertically scale the waveform data:
1. Move the cursors to specify the edit area to scale.
2. Push Operation (bottom)
!Vertical Scale... (pop-up)!OK (side).
3. Push the Factor side button.
This is the value by which you want to multiply the edit area waveform data.
Specify the scale using the general purpose knob or the numeric buttons. A
negative value of –100 to –1.01 inverts and rescales the signal. A value from
–1 to –0.01 inverts and reduces signal vertical values.
4. Push the Origin side button.
Specify the center of scale using the general purpose knob or the numeric
buttons.
5. Push the Exec side button.
The cursor-to-cursor data vertically expands or shrinks with the center at the
Origin position.
The Horizontal Invert... command horizontally inverts (flips) the edit area
waveform and marker data. You can invert the waveform and marker data
separately. This command does not change the waveform data record length.
Do the following steps to horizontally invert the waveform or marker data:
1. Move the cursors to specify the edit area to invert.
2. Push Operation (bottom)
!Horizontal Invert... (pop-up)!OK (side).
3. Push the Data, Marker1, or Marker2 side button to specify which data to
invert.
4. Push the Exec side button.
The data in the edit area inverts (flips) horizontally.
The Vertical Invert... command vertically inverts (flips) the edit area waveform
and marker data. You can invert the waveform and marker data separately. This
command does not change the waveform data record length.
Do the following steps to vertically invert the waveform or marker data:
1. Move the cursors to specify the edit area to invert.
2. Push Operation (bottom)
!Vertical Invert... (pop-up)!OK (side).
Horizontal Invert...
Vertical Invert...