Appendix B: Performance Verification

B-46

AWG610 Arbitrary Waveform Generator User Manual

Horizontal

Sweep . . . . . . . . . . . . . . . . . . . . . 0.5 ms/div

Trigger

Sou rce . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . DC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . +100 mV
Mode . . . . . . . . . . . . . . . . . . . . . . Au to

4. Set all the switches of the ground closure to open.

5. Follow the substeps below to set the AWG610 Arbitrary Waveform

Generator controls and select the sequence file:

a. Push UTILITY (front-panel)

!

System (bottom)

!Factory Reset (side)!OK (side).

b. Load the PT_EVENT.SEQ file.

Refer to Loading Files on page B–10 for file loading procedures.

c.

Push SETUP (front-panel)

!Run Mode (bottom)!Enhanced (side) to

set the enhanced mode.

6. Push the RUN and CH1 OUT buttons.

The LEDs above the RUN button and CH1 output connector are on.

7. Check the EVENT IN connector pin 0 input by doing the following

substeps:

a. Verify that the waveform being displayed on the oscilloscope is about

the same amplitude as shown in Figure B–21.