FG Mode

AWG610 Arbitrary Waveform Generator User Manual

3-225

Amplitude

The amplitude output voltage range is from 0.020 V

p-p

to 2.000 V

p-p

, in 1 mV

increments, terminated into a 50

W load. Set the waveform amplitude using the

LEVEL / SCALE knob, the numeric buttons or the general purpose knob.

Offset

The offset range is from –1.000 V to +1.000 V, in 1 mV increments. Use the
VERTICAL OFFSET knob, the numeric buttons or the general purpose knob to
set the waveform offset level.

Offset is also used for setup of DC level.

Polarity

This menu sets the output waveform polarity. Pushing the Polarity menu button
toggles polarity between Normal and Inverted.

Duty

When you select Pulse waveform, Duty... side menu is added. The Duty cycle is
set from 0.1% to 99.9% using the numeric buttons or the general purpose knob.
Increment step size depends on the output frequency. Refer to Table 3–56 on
page 3–227.