background image

The Setup Menu Screen

3-34

AWG610 Arbitrary Waveform Generator User Manual

This button lets you set the marker 1 and 2 high and low levels. You may set any
value from –1.1 to 3.0 V in 50 mV increments. The value of Low must always
be less than or equal to the value of High. The maximum marker output level
(High – Low) is 2.5 V

p-p

when the output signal is terminated into 50

W.

Do the following steps to set the marker signal high and low output levels:

1. Push SETUP (front)

!Vertical (bottom)!Marker... (side). The instrument

highlights the Marker screen icon for the selected channel.

2. Push the side menu button for the marker signal value you want to change.

3. Use the general purpose knob, numeric buttons, or keyboard to change the

marker value. If you use the general purpose knob, use the

a or ' buttons or

keys to select the digit to change.

This button selects the Normal or Direct output mode. In the Direct mode, the
instrument analog output is connected directly to the active channel output
connector, bypassing the internal analog filter and offset circuit. The analog
waveform signal output voltage range is from 20 mV

p-p

to 1.0 V

p-p

, in 1 mV

increments, terminated into 50

W. The maximum signal level is

±

0.5 V.

The filter setting and offset setting are not provided in the direct mode.

To connect the instrument analog output directly to the channel 1 and channel 2
output connectors, push SETUP (front)

!Vertical (bottom)!Output Mode

(side) to select Direct.

To connect the instrument analog output back to the signal vertical parameter
functions, push SETUP (front)

!Vertical (bottom)!Output Mode (side) to

select Normal.

The Horizontal Menu

The Horizontal menu lets you set waveform (analog and markers) horizontal
parameters for all output channels. The horizontal parameters include sample
clock source (internal or external), clock frequency, and marker signal delay
value. The Horizontal menu commands are Clock, Clock Ref, Marker 1 Delay,
and Marker 2 Delay.

The instrument uses only one clock sample frequency rate for all output signals,
regardless of individual waveform settings.

Marker...

Output