Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-35
3. Push UTILITY (front-panel)System (bottom)
!Factory Reset (side)!OK
(side).
4. Load the PULSE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
5. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
6. Verify the pulse response at 1 V amplitude by following the substeps below:
a. Verify that the rise time of the waveform displayed on the oscilloscope
from 10% to 90% point is equal to or less than 750 ps.
b. Set the oscilloscope sweep to 2 ns/div.
c.
Verify that the aberration of the displayed waveform on the oscilloscope
screen is within
±
10 %.
d. Set the oscilloscope sweep to 200 ns/div.
e.
Verify that the flatness of the displayed waveform on the oscilloscope is
within
±
3 % after 20 ns from the rising edge.
7. Do the following substeps to change the connection to check the CH1:
a. Push the CH1 output button.
b. Disconnect the cable from the CH1 output connector.
c.
Connect the cable to the CH1 output connector.
d. Push the CH1 output button to turn on the CH1 LED.
8. Repeat the Check Pulse Response procedure for the AWG610 Arbitrary
Waveform Generator CH1.
9. Push the CH1 output button to turn off the CH1 LED.
10. Disconnect the oscilloscope.