Tektronix AWG610 User Manual
Page 283
The UTILITY Window
AWG610 Arbitrary Waveform Generator User Manual
3-165
The calibration must be performed in the following cases:
H After a 20-minute warm up period
H Prior to high precision waveform output
H When the ambient temperature has changed more than +5 °C or less than
–5
°C from the previous calibration
Refer to the calibration and diagnostic screen to see if calibration has recently
been performed on the instrument. See Figure 3–56.
The calibration has completed when Done is displayed in the Calibration result
field. No calibration has been performed if the - - - is displayed. The factory reset
also causes the - - - to be displayed.
Figure 3-55: Calibration and diagnostic screen
NOTE. The calibration data in the memory may be lost if the instrument is
powered off while the calibrations are running.