10 mhz reference output levelătests – Tektronix AWG610 User Manual
Page 416
Appendix B: Performance Verification
B-56
AWG610 Arbitrary Waveform Generator User Manual
10 MHz Reference Output LevelĂTests
These procedures check the AWG610 Arbitrary Waveform Generator clock
output signal.
Equipment
required
A 50ĂW BNC coaxial cable and an oscilloscope.
Prerequisites
The instrument must meet the prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use the 50
W
BNC coaxial cable to connect the AWG610 Arbitrary
Waveform Generator 10 MHz REF OUT output to the oscilloscope input
connector (see Figure B–29).
10 MHz REF OUT
AWG610 Arbitrary Waveform
Generator rear panel
OscilloscopeĂ(TDS700)
50 W BNCĂcoaxial cable
Figure B-30: 10MHz Reference output level initial test hookup
2. Set the oscilloscope controls as follows:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . 0.2 V/div
CH1 offset . . . . . . . . . . . . . . . . . . 0 V
CH1 input impedance . . . . . . . . . . 50 W
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . 25 ns/div
Trigger
Source . . . . . . . . . . . . . . . . . . . . . CH1
Coupling . . . . . . . . . . . . . . . . . . . . AC
Slope . . . . . . . . . . . . . . . . . . . . . . Positive
Level . . . . . . . . . . . . . . . . . . . . . . 0 V
Mode . . . . . . . . . . . . . . . . . . . . . . Auto