AWG610 Arbitrary Waveform Generator User Manual

Table 3-12: Set Pattern dialog box parameters (cont.)

Parameter

Description

Pattern

Specifiesthe pattern field value. Enter the pattern data by

using the 0 or 1 numeric buttons on the front panel or from an

attached keyboard.
Push the Clear Pattern side button to clear the pattern data

field.
Push the Import Pattern side button to insert the edit area

pattern data from the active window target data type into the

pattern field. You can then write the pattern data to a target

waveform type in the active window or another window.

Target

Specifiesthe location where the generated data iscreated or

the source for imported pattern data. If you specify Data, the

pattern data replacesedit area waveform data; if Marker1 or

Marker2 isselected, the pattern data replacesthe edit area

marker data. To import the pattern from the Target specified

here, use the Import Pattern side menu.

Do the following steps to specify a pattern:

1. Move the cursors to specify the edit area in which to replace the waveform

data with pattern data.

2. Push Operation (bottom)

!Set Pattern... (pop-up)!OK (side).

The Set Pattern dialog box appears.

3. Select Data, Marker1, or Marker2 to specify the target data type to replace

with the pattern data.

4. Define the pattern using numeric buttons, or push the Import Pattern side

button to import the pattern data.

5. If necessary, you can change the pattern value by moving the cursor with the

a or ' button and then using numeric keys and the

key.

6. Push the OK side button to replace the waveform or marker data with the

specified pattern data.

The Import Pattern function lets you read waveform or pattern data from the
specified target data type of the active window and stores it in the pattern buffer.
You can then replace waveform or marker data with the pattern data in the
current window or another window. The Set Pattern dialog box converts all
waveform data greater than 0.5 volts to a one level if the waveform data is
analog T. All waveform data less than or equal to 0.5 volts is set to a zero level.