Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-23
Waveform
output
Gate
signal
CH2
CH1
Figure B-8: Relationship between gate signal and waveform output
d. Push SETUP (front-panel)
!Trigger (bottom)!Negative (side).
This changes the AWG610 Arbitrary Waveform Generator trigger polarity to
negative.
5. Verify that the oscilloscope displays a sine wave while the function generator
gate signal amplitude is Low level.
6. Turn off the function generator output and disconnect from the oscilloscope.