5.9

SCREEN MESSAGE

REASON FOR

MESSAGE

CORRECTIVE ACTION

M

ISCELLANEOUS

“Unable to Add -

Maximum number have
been entered”

Displays when adding a
control via QC Settings and
the maximum number of
controls (42), have already
been defined.

Press ok to return to the picklist of
existing controls to allow deletion of
one or more controls.

“Lot expired”

Expired cartridge lot was
entered via Test Settings,
QC Test, or Patient Test.

Use an in-date cartridge lot.

D

EVICE

S

ET

U

PDATE

“Unable to Transfer - Low

Battery”

Insufficient battery power to
begin device update.

Press quit to power-off analyzer;
replace the battery with a freshly
charged battery.

“Communication Error-

Verify connections and
receiving system”

Data transfer between the
analyzer and receiving
system was unsuccessful.

1. Verify that receiving system is ready

(e.g., DeviceCom is running).

2. Verify that cables and adapters are

properly connected to both the
IRMA TruPoint and receiving
system. Use cables and adapters
supplied by your service provider.

3. Verify that transfer cable is

connected to correct com port
(specified in DeviceCom setup).

4. If transferring serially, ensure that

the IRMA TruPoint and receiving
system baud rates match. If the
analyzer is configured to transfer to
“idms”, the receiving system must
be set to a baud rate of 19200. If the
analyzer is configured to transfer to
“host”, the receiving system must
be set to a baud rate of 9600.

5. If transferring by the LAN 10/100,

make sure that the network settings
are correct. See Section 7-
Communications
.