ITC IRMA TruPoint Blood Analysis System User Manual
Page 47

2.17
Patient Status
• Highlight the appropriate Patient Status setting.
› If Code or Room Air status is selected,
press
next
to accept the selected oxygen
status and complete entry.
› If Pump status is selected, press
next
to access the FIO
2
screen. Enter
FIO
2
value and press
next
to access Comments screen. Enter a one- to
twelve-character comment (optional) and press
next
to complete entry.
› If the Oxygen status is selected, press
next
to access the Oxygen
Therapy screen.
Oxygen Therapy
• Highlight the appropriate Oxygen Therapy
mode. Press
next
.
• The entry options displayed depend on the
relevant settings for the selected oxygen
therapy mode.
Ventilator Mode
• The
Ventilator Mode screen appears following
selection of Vent. from the Oxygen Therapy
screen. Highlight the appropriate mode and
press
next
.
• The following information may be entered, depending upon the ventilation
mode selected:
› Tidal Volume (mL)
› Rate or Frequency (B/min.)
› FIO
2
(%)
› PEEP (cm H
2
O)
› IPAP (cm H
2
O)
› EPAP (cm H
2
O)
› Comments (up to twelve alpha-numeric characters)
Mask Type
• The Mask Type screen appears following
selection of Mask from the Oxygen Therapy
screen. Highlight the appropriate mask type and
press
next
.