CARTRIDGE/ANALYTE CONFIGURATIONS

IRMA TruPoint cartridges are available in the following analyte con-
figurations:

Table B-6

Cartridge

Measured

Calculated

BG

pH,

pCO

2

,

pO

2

HCO

3

-

, TC

O

2

, BEb, BEecf,

O

2

Sat

CC

pH,

pCO

2

,

pO

2

, Hct,

Na+

,

K+

, iCa

HCO

3

-

, TC

O

2

, BEb, BEecf,

O

2

Sat, tHb, iCa(N)

H3

Hct,

Na+

,

K+

, iCa

tHb

H4

Hct,

Na+

,

K+

, Cl-,

BUN/urea

tHb

GL

Glu,

Na+

,

K+

, Cl-

CR

Creatinine

MDRD

GFR

LA

Lactate

CARTRIDGE EQUILIBRATION TIMES

Following removal from their shipping container, IRMA TruPoint cartridges
must equilibrate to their storage environment prior to use (see Section 1-
Cartridge Storage and Equilibration Procedure). Equilibration times depend
on the product type as follows:

Table B-7

Cartridge Equilibration

Time (Hours)

Storage Temperature Warm-up

Time

(Minutes)

BG

72

15 - 30 °C (59 - 86°F)

None

CC

72

15 - 30 °C (59 - 86°F)

None

H3

1

15 - 30 °C (59 - 86°F)

None

H4

1

15 - 30 °C (59 - 86°F)

None

GL

1

15 - 30 °C (59 - 86°F)

None

CR

None

2 - 8 °C (35.6 – 46.4°F)

15

LA

1

15 - 30 °C (59 - 86°F)

None

Refrigerated cartridges must be removed from the refrigerator and sit at room
temperature before they can be used (Table B-7).