ITC IRMA TruPoint Blood Analysis System User Manual
Page 57

3.5
a. Passed EQC Test:
Manual
a1. Press
view
to view quantitative results
(optional). Press
next
from each screen to
proceed through results for all analytes.
a2. Press
notes
to select QC notes to attach to the QC record (optional).
a3. Press
to obtain a printout. (Results may automatically print,
depending on auto-print setup).
a4. Press
done
to return to the QC TEST OPTIONS menu.
Automatic
a5. Results will print (if the printer is configured to print) and the
analyzer will display the main menu or user
ID screen (if required).
a6. If printer is not configured to print, there
will be no print out. The main MENU will
be displayed.
b. Failed EQC Test:
Manual
b1. Press
help
for assistance in resolving the problem that caused the
failure (optional).
b2. Press
notes
to select QC notes to attach to the record (optional).
b3. Press
to obtain a printout. (Results may automatically print,
depending on auto-print setup).
b4. Press
done
to return to the QC TEST OPTIONS menu.
Automatic
b5. If printer is configured to print, “EQC failed” will print and the “EQC
Results failed” screen will display.
b6. Press
done
to return to the main MENU or the User ID entry screen (if
required). Another auto-EQC test will initiate following the next
inactivity period, or a manual EQC test can be performed.