ITC IRMA TruPoint Blood Analysis System User Manual
Page 134

B.2
General Specifications, continued
Table B-1
Category Specification
Display:
Liquid Crystal Display (LCD) touch
screen
Power Requirements:
Cartridges: none
Analyzer:
7.2 V, 2 amp by rechargeable battery or
AC adapter
Battery Charger/Power Supply:
100 - 240 Vac, 1 Amp, 50 - 60 Hz; The
battery charger is a Class 1LED
Product
AC Adapter:
100 - 240 Vac, 1.0 Amps, 50 - 60 Hz
Empty Battery Recharge
Cycle:
2.5-7 hours (NiMH)
Analyzer and AC Adapter Classifications:
When used at the bedside either with battery or AC Adapter,
Equipment Classifications As Defined Per IEC60601-1:2003/
IEC60601-1 2nd Edition
• Protection Against Electrical Shock: Class II, Internally Powered
Equipment with no applied parts
• Protection Against Ingress of Liquids: Ordinary (no protection as defined
by IEC 529)
• Product Cleaning and Disinfection: Only according to recommendations
of the manufacturer’s accompanying documentation
• Mode of Operation Of Equipment: Short-time
• Degree of Safety of Application in the Presence of Flammable
Anesthetic Mixture With Air, Oxygen or Nitrous Oxide: Not Suitable
Note:
• as defined in the above standards, the classification of Not Suitable is not
intended to indicate that the instrument is not suitable for use in an
Operating Room (OR) environment. Rather, it is intended to indicate that
the instrument is not suitable for use in the direct presence of a
flammable anesthetic mixture with air, oxygen or nitrous oxide.
Device Disposal (at end-of useful life):
The analyzer and accessories may be returned to ITC.
Nickel Metal Hydride (NiMH) batteries are classified as a non-hazardous
waste and should be disposed of in accordance with the local, state, or
federal regulations where they are used. Contact your local government for
disposal practices in your area. Although NiMH batteries are considered
environmentally friendly, they may also be recycled. Recycling options
available in your local area should be considered when disposing of this
product or batteries may be returned to ITC for recycling.