beautypg.com

ITC IRMA TruPoint Blood Analysis System User Manual

Page 64

background image

3.12

22. If a sensor errors during the analysis phase of a liquid QC test, the

result for that sensor and any other sensors that are dependent on that
sensor will be suppressed. Suppressed results will be dashed-out on
both the screen and printout, and will not satisfy QC lockout
requirements for the associated analyte group (with or without QC
limits established).

Document QC Notes (optional)

23. The QC Notes feature allows pre-defined notes to be permanently

associated with individual QC test records. Up to 3 notes can be
attached to an individual QC test record. If the QA User enabled QC
Notes, a notes button will be present on the Results screen. Press the

notes

button to access the pre-defined notes available for selection.

Use the up and down arrows to highlight the desired note and press
the check box button. Repeat for additional notes, and press

next

when complete.

Obtain a Printout (optional)

24. a. If the Auto-print setting is “immediate”, test results will

automatically print when the analysis is complete.

b. If the Auto-print setting is “delayed”, test results will automatically

print when the

done

button is pressed from the QC Test Results

screen.

c. If the Auto-print setting is “off”, a printout may be obtained by

pressing the

print

button.