22. If a sensor errors during the analysis phase of a liquid QC test, the
result for that sensor and any other sensors that are dependent on that
sensor will be suppressed. Suppressed results will be dashed-out on
both the screen and printout, and will not satisfy QC lockout
requirements for the associated analyte group (with or without QC
limits established).
Document QC Notes (optional)
23. The QC Notes feature allows pre-defined notes to be permanently
associated with individual QC test records. Up to 3 notes can be
attached to an individual QC test record. If the QA User enabled QC
Notes, a notes button will be present on the Results screen. Press the
notes
button to access the pre-defined notes available for selection.
Use the up and down arrows to highlight the desired note and press
the check box button. Repeat for additional notes, and press
next
when complete.
Obtain a Printout (optional)
24. a. If the Auto-print setting is “immediate”, test results will
automatically print when the analysis is complete.
b. If the Auto-print setting is “delayed”, test results will automatically
print when the
done
button is pressed from the QC Test Results
screen.
c. If the Auto-print setting is “off”, a printout may be obtained by
pressing the
button.