2.9

b2. To enter a new cartridge Lot for the

first time, highlight

new

from the

picklist and press

edit

.

› Enter the cartridge Lot using the

alpha keypad. Press

next

.

› Enter the Cal Code using the alpha

keypad. Press

next

to return to the

Verify Information screen. If the
information is correct, press

next

to

continue patient test. If the
information is incorrect, press

edit

to re-enter the Lot and Cal Code.

Note:

• IRMA TruPoint generates an error if the calibration code is not verified

within approximately 2 minutes of cartridge insertion.

Select Analytes

13. If the QA User set up the analyzer to always automatically report results

for a specified group of analytes, the Select Analytes screen will not
appear; the analytes defined in the setup will always automatically be
reported. Proceed to step 16.

14. If the QA User set up the User Selects

option for the product being run, the Select
Analytes
screen will appear with each test.

15. Select the analytes or analyte groups to be tested by highlighting the

appropriate button(s). Press

next

when all desired analytes have been

selected.

Note:

• Analytes that are locked-out will have a

padlock over the selection button, preventing
selection of that analyte until QC Lockout
requirements have been satisfied.

Calibration

16. Once the required data has been entered or verified, the calibration

process begins.