ITC IRMA TruPoint Blood Analysis System User Manual
Page 41
2.11
Note:
• If air bubbles or calibration gel are present in the sample path after the
initial sample injection, remove them by injecting additional sample from
the same syringe, or by lightly tapping the top of the syringe plunger and
then injecting additional sample from the same syringe.
• If air bubbles or calibration gel cannot be removed from the sample path,
press
cancel
to stop the test, discard the cartridge, and begin again with a
new cartridge.
Proper injection technique will prevent air bubbles or calibration
gel from being present in the sample path after initial sample injection.
20. Ensure that no air bubbles or calibration gel are present in the sample
path, then immediately press
test
to continue sample analysis.
a. The Analyzing screen displays if there is no
required patient bypass status or patient ID
information to be entered.
b. If patient bypass status or patient ID
information entry is required, but not yet
entered, the Bypass Status screen or
Patient ID screen will display when
test
is
pressed. The required information must be
entered
before results will display or print. Refer to
Section 2-Entering Test Information and
Section 2-Patient Bypass Status for details.
Enter the required information and press
next
to advance to the Analyzing (or
Results) screen.
21. When the test is complete, remove the cartridge with the collection
device attached. Dispose of both in accordance with established
guidelines for your facility.
View Test Results
22. Results automatically appear on the IRMA TruPoint touchscreen when
the analysis is complete. Press
more
to view additional measured or
calculated results if all the results do not appear on the screen.