ITC IRMA TruPoint Blood Analysis System User Manual
Page 155

D.3
ACCURACY - CONTINUED
Glucose (Glu)
Number of samples:
37
Range evaluated (mg/dL):
19 -338
Slope:
0.970
Intercept:
5.47
Correlation Coefficient (r):
0.992
Sy.x:
11.08
Creatinine (Cr)
Number of samples:
110
Range evaluated (mg/dL):
0.5 - 10.1
Slope:
0.940
Intercept:
0.16
Correlation Coefficient (r):
0.981
Sy.x:
0.55
Lactate (LA)
Number of samples:
117
Range evaluated (mmol/L):
1.1-11.5
Slope:
0.959
Intercept:
-0.032
Correlation Coefficient (r):
0.98
Sy.x:
0.46