ITC IRMA TruPoint Blood Analysis System User Manual
Page 116
7.20
>
QA Users
have access to all analyzer setting and testing
functions, including User ID settings.
>
General Users
may perform and recall all tests, (as configured
by QA User), but have limited access to analyzer set-up
functions.
>
Gluc Strip Users
may perform and recall only glucose strip,
EQC, and Temperature tests (if applicable), and have limited
access to analyzer set-up functions.
Edit User:
Privileges can be changed for existing User IDs.
Delete User:
Existing User IDs can be deleted from analyzer memory.
Print Lists:
A list of all User IDs, grouped by Privileges category, can be
printed.
Note:
• The
Gluc Strip User privilege is available for users that have purchased
the SureStepPro Glucose Module, an optional side attachment to the
IRMA TruPoint analyzer. Refer to the SureStepPro Glucose Module
User Manual for details.
• The
General User privilege allows both IRMA TruPoint cartridge and
glucose strip testing (if glucose module is present).
DEFAULT QA USER ID
The IRMA TruPoint analyzer is shipped with a default QA User ID of
123456. Use the default QA User ID to access the User ID Options screen
and enter a new QA User ID(s). Add new QA
Users before deleting the default QA User ID.
ADDING A NEW QA USER ID
1. Press
User ID
from the SETTING OPTIONS
menu.
2. Highlight
Add User
from the picklist on the
User ID Options screen. Press
edit
.
3. Enter the new QA User ID. Press
next
.
4. Highlight button next to
QA
on the User
Privileges screen. Press
next
.