ITC IRMA TruPoint Blood Analysis System User Manual

Page 38

background image

2.8

Do not touch the cartridge leads
after the tape has been removed.
Do NOT remove the luer cap
from the cartridge. (Figure 2-4).

10. Fully insert the cartridge into the

analyzer within 15 minutes of
opening the package. (Figure 2-5)

Note:

• Do not use a cartridge if it has

been out of the package longer
than 15 minutes; do not reuse a
single-use cartridge once it has
been inserted into the analyzer.

Figure 2-5

Verify/Enter Cartridge Information

11. Ensure that the product type (Type)

displayed on the Verify Information screen
is correct. If the product type displayed does
not match the type of the cartridge inserted
into the analyzer, press

back

to return to the

Select Product Type screen.

12. The Verify Information screen displays the cartridge Lot and Cal Code

from the most recent test performed on the same product type.

a. If the Lot displayed matches the Lot of the cartridge inserted into the

analyzer, press

next

. Proceed to step 13.

b. If the Lot displayed does not match the Lot of the cartridge inserted

into the analyzer, press

edit

to choose a different previously-entered

Lot, or enter a new Lot for the first time.

b1. To choose a previously-entered Lot,

highlight the correct cartridge Lot
from the picklist. Press

next

to return

to the Verify Information screen. If the
information is correct, press

next

to

continue patient test. If the information is incorrect, press

edit

to

re-enter the Lot and Cal Code.