D.2
ACCURACY - CONTINUED
Potassium (K+)
Number of samples:
137
Range evaluated (mM):
1.5 - 13.8
Slope:
0.961
Intercept:
-0.033
Correlation Coefficient (r):
0.999
Sy.x:
0.217
Ionized Calcium (iCa)
Number of samples:
137
Range evaluated (mM):
0.80 - 5.70
Slope:
1.019
Intercept:
-0.063
Correlation Coefficient (r):
0.997
Sy.x:
0.136
Chloride (Cl-)
Number of samples:
56
Range evaluated (mM):
74.8 - 136.4
Slope:
0.956
Intercept:
4.64
Correlation Coefficient (r):
0.986
Sy.x:
3.0
Hematocrit (Hct)
Number of samples:
137
Range evaluated (%):
22.0 - 60.5
Slope:
0.883
Intercept:
2.645
Correlation Coefficient (r):
0.978
Sy.x:
2.45
Blood Urea Nitrogen (BUN)
Number of samples:
56
Range evaluated (mg/dL):
12.5 - 74.6
Slope:
0.985
Intercept:
0.46
Correlation Coefficient (r):
0.995
Sy.x:
1.9