beautypg.com

Electronic quality control – ITC IRMA TruPoint Blood Analysis System User Manual

Page 56

background image

3.4

ELECTRONIC QUALITY CONTROL

Electronic Quality Control tests may be performed two ways:

1. Manual EQC Test: initiated by user by pressing the

EQC

button on the

QC Test Options screen.

2. Automatic EQC Test: runs automatically per a pre-configured schedule.

This feature requires the use of an AC adapter to power the IRMA
TruPoint.

PERFORMING A MANUAL EQC TEST

Turn IRMA TruPoint Analyzer On

1. Touch the right-hand edge of the screen to turn the analyzer “on”.

Enter User ID (optional)

2. If the User ID option has not been activated, go to step 4.

3. If the User ID option is activated, the Enter or

Scan ID screen will display. The User ID may
be entered using either the alpha/numeric
keypads or the bar code scanner.

a. Entry via touchscreen keypads - enter your

User ID and press

next

.

b. Entry via bar code scanner - scan your User ID. The IRMA TruPoint

will automatically advance to the next screen.

Initiate an EQC Test

4. Press

QC Test

on the main MENU. The QC

TEST OPTIONS menu displays.

5. Press

EQC

on the QC TEST OPTIONS menu

to initiate the EQC test.

6. The

Analyzing screen appears, indicating that

the diagnostic tests are in progress.

Review Results

7. Once the EQC test is complete, the analyzer

displays the EQC Results status screen. The
initial result for an EQC test is either “Passed”
or “Failed”.